Edwards nEXT 240D - NEW

Starting at $13,195
In stock
Only %1 left
SKU
NEW_nEXT240D
Weight 12.5 lbs
Pumping Speed 240 l/s
Ultimate Pressure < 6 x 10-8 mbar

Edwards nEXT240D Turbomolecular Vacuum Pump - NEW

The nEXT240D is a hybrid bearing compound turbo pump available with either an ISO-100 or a CF-100 inlet flange and has a N2 pumping speed of 240 l/s. nEXT pumps combine our proven bearing technology (oil lubricated ceramic lower bearing with dry permanent magnetic upper bearing), an improved rotor design with a new molecular drag stage to deliver improved pumping speed and compression ratios, and true user serviceability. They feature 24V to 48 VDC sensor less motors with a built in drive that is fully compatible with our range of TAG and TIC controllers. They are available pre-set for either 80W or 160W maximum power, the former enabling use with our 100W TIC controllers are also recommended for pumping Argon, but with a longer ramp time to full operating speed. A basic pump system would require a nEXT240D pump, a TAG or TIC controller, either an air or water cooling accessory and a suitable backing pump.

  • Value for money - exceptional pumping speeds and compression ratios
  • No unplanned downtime - field proven reliability in the most demanding environments
  • Easy integration - standard fit in many scientific instruments
  • Reliability for peace of mind - over 200,000 turbomolecular pumps installed
  • Low cost of ownership - end user serviceable

Questions? Please call our toll free number at 866-332-0500 or send an email to our sales team!

Inlet flange DN 100 ISO-K / DN 100 CF
Outlet flange DN 25 NW
Pumping speed  
N2 240 l/s
He 230 l/s
H2 165 l/s
Compression ratio D / T  
N2 > 1 × 1011
He 3 x 105 / 1 x 106
H2 1 x 104 / 1.5 x 104
Ultimate pressure with RV backing pump ISO / CF < 6 x 10-8 / < 5 x 10-10 mbar
Recommended backing pump* RV12 / nXDS10i
Vent port 1/8 inch BSP female
Purge port 1/8 inch BSP female
Max continuous inlet flow (Nitrogen)†  
Water cooling (water at 15°C, ambient temp at 40°C) 45 sccm
Forced air cooled, 35°C ambient 30 sccm
Pump rotational speed  
Nominal rotational speed 60,000 rpm
Standby rotational speed Variable from 33,000 to 60,000 rpm (42,000 rpm default)
Programmable power limit settings Variable from 50 - 200 W (160W default)
Start time to 90% speed D / T 160 W 115 / 140
Analogue outputs Pump rotational speed;
Power consumption; Pump temp; Controller temp
Cooling method Forced air / water
Ambient air temp for forced air cooling 5 - 35°C
Min cooling water flow rate (water 15°C) 15 l/h-1
Water temp range 10 - 20°C
Max inlet flange temp 75°C
Operating attitude Vertical and upright, through to Horizontal +/- 2°
Noise level at 1 metre < 45 dB(A)
Max magnetic field pump can tolerate 5 mT
Recommended controller TIC 200 turbo and instrument controller
Quiescent electrical power Typically 20W

Specifications

Need modifications to this item? PTB Sales' expert team of engineers can configure products to your specific needs.

Edwards MODEL: nEXT240 Turbomolecular Vacuum Pump - NEW

CODE: B812-00-100 / B81200100  nEXT240D  ISO-100 160W 
CODE: B812-00-200 / B81200200  nEXT240D  CF-100 160W
CODE: B812-00-101 / B81200101  nEXT240D  ISO-100 80W
      B812-00-201 / B81200201  nEXT240D  CF-100 80W
      B813-00-100 / B81300100  nEXT240T HC  ISO-100 160W
      B813-00-101 / B81300101  nEXT240T HC  ISO-100 80W
      B813-00-200 / B81300200  nEXT240T HC  CF-100 160W
      B813-00-201 / B81300201  nEXT240T HC  CF-100 80W